Add time:07/24/2019 Source:sciencedirect.com
Phosphate selective electrodes have been produced based upon rubbery membranes containing heterocylic macrocycles as sensors covalently bound to a cross-linked polystyrene-block–polybutadiene-block–polystyrene (SBS) polymer. The membranes were robust and the best HPO42−-selective membrane fabricated was composed of 7.1% (m/m) dicumyl peroxide, 28.3% (m/m) 2-nitrophenyloctylether, 9.8% (m/m) 3-(10-undecenyl)-1,5,8-triazacyclodecane-2,4-dione, 31.0% (m/m) SBS and 23.8% (m/m) PoleStar™ 200R (clay-based filler). The characteristics of this electrode were a linear Nernstian range of 3.9×10−3 to 1×10−6 mol dm−3 HPO42− with a limit of detection of 1.0×10−6 mol dm−3 HPO42−, a slope of −29.7±0.9 mV per activity decade and a pH range from 6 to 8. Selectivity coefficients for phosphate against various interfering anions (chloride, sulfate and nitrate) were determined. Response times were 2 min or under, stability of response and electrode lifetime in continuous use were also very satisfactory. The response behavior of HPO42−-ISEs based upon mobile and bound ionophores was comparable and suggests that mobility of the ionophore is not necessary to obtain a working ISE and that covalent binding of ionophores can be used to produce ISEs of increased stability and robustness.
We also recommend Trading Suppliers and Manufacturers of dipotassium octyl phosphate (cas 19045-79-5). Pls Click Website Link as below: cas 19045-79-5 suppliers