Add time:07/25/2019 Source:sciencedirect.com
A chromatographic method (high-performance liquid chromatography/photodiode array detection) for the simultaneous determination of itraconazole and its mayor metabolite, HYDROXY ITRACONAZOLE (cas 112559-91-8), was developed. The validation data demonstrated acceptable values for linearity (dynamic range between 0.25–16 μg/mL for hydroxy itraconazole, 0.125–4 μg/mL for itraconazole), precision (Coefficient of Variation, 2.23–4.28% and 1.61–9.52% respectively), accuracy (Relative Error, −2.89 to −17.26% and −3.11 to 14.44%), selectivity, sensitivity (Lower limit of Quantification, 0.125 and 0.25 μg/mL), and recovery (97.30–105.82%) for both compounds. Its clinical suitability was also investigated in 104 trough human serum samples. The assay resulted to be rapid, sensitive, and simple for simultaneously quantification of these azole compounds in serum samples and should make a positive contribution to optimize therapies on an individual basis.
We also recommend Trading Suppliers and Manufacturers of HYDROXY ITRACONAZOLE (cas 112559-91-8). Pls Click Website Link as below: cas 112559-91-8 suppliers