Add time:08/14/2019 Source:sciencedirect.com
This paper proposes a very simple procedure for preparing a biocompatible sensor based on a protein (bovine serum albumin, BSA), enzyme and Vinylferrocene (cas 1271-51-8) (VF) composite membrane modified electrode. The membrane was prepared simply by first casting vinylferrocene and then coating it with BSA and glucose oxidase immobilised with glutaraldehyde. The sensor response was independent of dissolved oxygen concentration from 3 to 10 ppm and showed good stability for serum sample measurement, unlike the commonly used BSA/enzyme modified electrode. The sensor response was almost unchanged over the measurement time (>10 h) whereas the responses of a BSA and glucose oxidase modified platinum electrode and an osmium-polyvinylpyrridine wired horseradish peroxidase modified electrode (Ohara et al., 1993) fell to 68% of their initial value in a serum sample containing 10 mM glucose.
We also recommend Trading Suppliers and Manufacturers of Vinylferrocene (cas 1271-51-8). Pls Click Website Link as below: cas 1271-51-8 suppliers