Encyclopedia

  • Online photocatalytic device for highly selective pre-column fluorescence derivatization of 5-hydroxyindoles with benzylamine
  • Add time:08/14/2019         Source:sciencedirect.com

    A fluorimetric liquid chromatographic method for the determination of 5-hydroxyindoles based on the benzylamine derivatization process mediated through an online photocatalytic oxidation has been developed. In this study, we used a photocatalytic column comprising tefzel tubing packed with TiO2-coated glass beads, as a pre-column derivatization reactor. The fluorescence derivatization of 5-hydroxyindoles using benzylamine proceeded during their passage through the reaction column under near-UV irradiation. The 5-hydroxyindole derivatives were separated continuously on a reversed-phase liquid chromatography within 50 min, using 100 mM acetate buffer (pH 4.6)–acetonitrile (72:28, v/v; isocratic elution) containing 3 mM sodium octanesulfonate; the samples were detected fluorimetrically at 465 nm upon excitation at 350 nm. The detection limits (signal-to-noise ratio = 3) of the 5-hydroxyindoles were in the range from 160 to 360 fmol per 5 μL injection. We have applied this method, which requires minimal sample pre-treatment, to the determination of 5-hydroxyindole-3-acetic acid in human urine.

    We also recommend Trading Suppliers and Manufacturers of 5-HYDROXYINDOLE-3-ACETONITRILE (cas 100929-86-0). Pls Click Website Link as below: cas 100929-86-0 suppliers


    Prev:Purification and some properties of cathepsin H from rabbit skeletal muscle
    Next: Concerted derivatization and concentration method with dispersive liquid–liquid microextraction for liquid chromatographic analysis of 5-hydroxyindoles in human serum)

About|Contact|Cas|Product Name|Molecular|Country|Encyclopedia

Message|New Cas|MSDS|Service|Advertisement|CAS DataBase|Article Data|Manufacturers | Chemical Catalog

©2008 LookChem.com,License: ICP

NO.:Zhejiang16009103

complaints:service@lookchem.com Desktop View