Add time:08/17/2019 Source:sciencedirect.com
A fractured Vitallium (cas 12629-02-6) plate used as a fixation device was investigated for determining the basic mechanism/s responsible for premature failure. Microhardness tests, tensile tests and extensive scanning electron microscopy were carried out to establish the dominant mechanism of fracture. From the detailed fracture surface analysis and the background information provided on the implant, it was determined that the implant failed by the mechanism of corrosion fatigue at the interface of implant plate and screw used in the bone fixation device.
We also recommend Trading Suppliers and Manufacturers of Vitallium (cas 12629-02-6). Pls Click Website Link as below: cas 12629-02-6 suppliers