Two types of isoelastic endosseous dental implants were produced and their functionality was tested. One type consisted of a porous sintered TiTaSO alloy, the other had a special surface structure consisting of titanium wire loops. Their mechanical properties were optimized by the production parameter (sintering and diffusion bonding, respectively). The functionality was tested after insertion into an artificial jaw which had properties corresponding to the natural mandibular. The elastic properties of both implants were similar to the properties of the bone. In addition the implants have a safe anchorage bone ingrowth. In animal experiments using the implant with surface loops it was observed that the bone entered the loops and even extremely small surface cavities in the wire loops.